Carefusion Alaris 8100 Features

The Carefusion Alaris Infusion Pump 8100 module is designed for the delivery of fluids, medications, blood, and blood products using continuous or intermittent delivery for adult, pediatric or neonatal patients. Infusion delivery can occur via clinically acceptable routes of administration (e.g., intravenous, intra-arterial, subcutaneous, epidural, enteral, or irrigation of fluid spaces).

Clinicians can attach up to four pump modules to a single Point of Care (PC) unit allowing four different infusions. The pump modules can be attached to either side of the PC unit to create the desired configuration.

The Alaris Pump module supports the Guardrails® Suite MX software. This software helps to reduce the risk of medication errors by providing a test of reasonableness before the initiation of therapy.

Carefusion Alaris 8100 Specifications

Dimensions

  • 3.3″W x 8.9″H x 5.5″D
  • Weight: 2.5 lbs

Environmental

Operating

  • Temperature Range: 41 to 104°F (5 to 40°C)
  • Relative Humidity: 20 to 90%
  • Atmospheric Pressure: 525 to 4560 mmHg (700 to 6080 hPa)

Storage / Transport

  • Temperature Range: -4 to 140°F (-20 to 60°C)
  • Relative Humidity: 5 to 85%
  • Atmospheric Pressure: 375 to 760 mmHg (500 to 1013 hPa)

General

  • Critical Volume: The maximum over-infusion which can occur in the event of a single fault condition is 0.6 mL
  • Fluid Ingress Protection: IPX1, Drip Proof Infusion of Air, Means to Protect Patient from Ultrasonic Air-in-Line Detection Maximum single bolus size = selectable 50, 75, or 250 microliters nominal (500 microliters in Anesthesia Mode)
  • Infusion Pressure, Maximum: 654 mmHg (Maximum Occlusion Alarm Threshold plus tolerance)
  • KVO (Keep Vein Open) Rate: Factory Default Setting is 1 mL/h if the set rate is 1 mL/h or above; or set rate if the rate is 0.9 mL/h or below
  • KVO Selection Range: KVO rate can be set in System Configuration from 0.1-20 mL/h in 0.1 mL/h increments
  • Occlusion Alarm Thresholds:
    • Pump Mode: 525 mmHg at rates ≥30 mL/h; Varying level based on rate and patient back–pressure at rates <30 mL/h
    • Selectable Mode: User selected from 50 to 525 mmHg in 25 mmHg increments
  • Operating Principle: Positive displacement Rate Accuracy: The rate accuracy of the Medley Medication Safety System is ±5% at rates between 1 and 999 mL/h and ±5.5% at rates <1 mL/h, 95% of the time with 95% confidence, under the conditions listed belowInfusion Rate Range: 0.1 to 999 mL/h
    Ambient Temperature: 68 ±4°F (20 ±2°C)
    Source Container Height: 20 inches above the top of the Pump Module
    Test Solution: Distilled Water
    Distal Back pressure: 0 mmHg (0 kPa)
    Needle: 18 gauge
    Administration Set Model 2210

 